INFICON Stripe CDG045Dhs 0.01 ... 1000Torr / mbar, new 10mTorr User Manual

Page 11

Advertising
background image

tina84e1 (2014-07)

11

Remote Zero Adjust


External switching contact

digital input for zero adjust-
ment (

→  24)

30 V (dc) / <5 mA (dc)

Switching functions

setpoints SP1/2, ATM, error
status

Setpoint relay

Setting range

1 … 99% F.S. (0.1 … 9.9 V)

Hysteresis

4)

1%

F.S.

Switching characteristics

4)

Low Trip Point (default)

Relay contact

30 V (dc) / ≤0.5 A (dc)
floating (NO)

closed p

≤ p

SP

(LED lit solid)

open p

≥ p

SP

(LED off)

Status relay

Relay contact

30 V (dc) / ≤0.5 A (dc)
connected to supply com-
mon (pin 5)

closed measurement

mode

warning

open no

supply voltage

warming up, error

Diagnostic port

Cable

Mini USB type B, 5-pin
USB cable type A/mini B

EtherCAT interface

Specification, data format,
communication protocol


→  [5], [6]

Data rate

100 Mbps

Note address

explicit device identification

Physical layer

100Base-Tx (IEEE 802.3)

4

)

The hysteresis and the switching characteristics can be programmed via

the serial interface or the diagnostic port.

Advertising