4.8 EtherCAT
Operation
Caution
Caution: data transmission errors
Any attempt to simultaneously operate the gauge
via the EtherCAT interface and the diagnostic port
causes data transmission errors.
Therefore the gauge must not be operated simul-
taneously via the EtherCAT interface and the diag-
nostic port.
For operating the gauge via EtherCAT, prior installation of the
device specific ESI file is required on the bus master side. This
file can be downloaded from our website.
Explicit Device Address Setting
(default
00
hex
)
During device initialization, the device address switches are read
by the device firmware. This device address is supported to the
master as Explicit Device Identification.
2 4
0
6
8
A
E
2 4
0
6
8
A
E
2 4
0
6
8
A
E
The explicit device address is set in hexa-
decimal form (00 … FFF
hex
) via the
<x100>, <x10> and <x1> switches.
Example: Device address = 0xDDD (dec 3549):
0x100 * 0xD (dec 3328) + 0x10 * 0xD (dec 208) + 0x1 * 0xD
(dec 13)
Status LED
LEDs on the gauge inform on the gauge status and the current
EtherCAT status (
→ [3]).