Push the <ZERO> button again:

Fine adjustment
within 0...3 s:

the zero adjustment value changes
by one unit (push <ZERO> button in
intervals of 1 s)

Change of direction
within 3...5 s:

the zero adjustment changes its
direction (the blinking frequency of
the <STATUS> LED changes
briefly)

If the <ZERO> button is released for more than 5 s, the

gauge returns to the measurement mode.

The <STATUS> LED blinks green if the signal output is negative
(< -20 mV).

4.3

Switching Functions SP1, SP2

The two switching functions can be set to any pressure within
the measurement range of the gauge. A solid state relay is
provided for each switching function.

The current threshold setting
• is output at the measurement signal output instead of the

pressure signal and can be measured with a voltmeter after
the <SP1> or <SP2> button is pressed

• can be read / written via the diagnostic port (→  [4]) and the

EtherCAT interface (

→  [3]).

Switching characteristics and hysteresis

The switching characteristics and the hysteresis of each

set point can only be programmed via the diagnostic
port (

→  [4]) and the EtherCAT interface (→  [3]).