Zero adjustment using the ramp function can be

performed via
• the <ZERO> button on the gauge,
• the EtherCAT interface (→  [3]),
• the diagnostic port (→  [4]),

Recommended procedure for adjusting the offset of a

measuring system:

→ Notice  24.

Operate the gauge for at least 2 hours (until the signal is

stable).

Push the <ZERO> button with a pin (max. ø1.1 mm) and

keep it depressed. The <STATUS> LED starts blinking.
After 5 s, the zero adjustment value, starting at the current
output value, keeps continually changing (ramp) until the
button is released or until the setting limit (max. 25% F.S.)
is reached. The corresponding output signal is delayed by
about 1 s.

max. ø1.1 mm

Keep the button
depressed