DANGER

DANGER: protective ground
Products that are not correctly connected to ground
can be extremely hazardous in the event of a fault.
Electrically connect the gauge to the grounded
vacuum chamber. This connection must conform to
the requirements of a protective connection ac-
cording to EN 61010:
• CF and VCR flanges fulfill this requirement.
• For gauges with a KF flange, use a conductive

metallic clamping ring.

• For gauges with a ½" tube, take appropriate

measures to fulfill this requirement.

Caution

Caution: vacuum component
Dirt and damages impair the function of the vac-
uum component.
When handling vacuum components, take appro-
priate measures to ensure cleanliness and prevent
damages.

Caution

Caution: dirt sensitive area
Touching the product or parts thereof with bare
hands increases the desorption rate.
Always wear clean, lint-free gloves and use clean
tools when working in this area.

Mount the gauge so that no vibrations occur.