6 activating the factory setting (factory reset) – INFICON Stripe CDG045Dhs 0.01 ... 1000Torr / mbar, new 10mTorr User Manual

Page 34

Advertising
background image

34

tina84e1 (2014-07)

The current ATM threshold setting
• can be read / written via the diagnostic port
• is output at the measurement signal output instead of the

pressure signal, can be measured with a voltmeter, and is
displayed on the LCD display after the <SP1> or <SP2>
button is pressed

• can be read / written via the EtherCAT interface.

DANGER

DANGER: malfunction
If processes are controlled via the signal output,
keep in mind that by pushing the <SP> button the
measurement signal is suppressed and the cor-
responding threshold value is output instead. This
can cause malfunctions.
Push the <SP> button only if you are sure that no
malfunction will cause.

Programming ATM setpoint

Programmable parameters:
(

→  [3], [4])

Factor of ATM
Low Trip Enable
Low Trip Point Hysteresis

High

Trip

Enable

High Trip Point Hysteresis

Setpoint

Mode


Switching characteristics of the setpoints

→  28.

Diagnostic port

→  [4].


4.6

Activating the Factory Setting (Factory Reset)

All user defined parameters (e.g. zero, filter) are restored to their
default values.