tina84e1 (2014-07)

33

4.5 ATM

Setpoint

The setpoints SP1 and SP2 of gauges with separate atmos-
pheric pressure sensor

8)

can be programmed to atmospheric

pressure setpoint (ATM setpoint) via the diagnostic port or the
EtherCAT interface.
The ATM setpoint is defined as a factor of the current atmos-
pheric pressure. The relay switches when the pressure in the
vacuum system has reached the defined value (differential
pressure in relation to the atmospheric pressure).

Example: ATM setpoint: 110% of the atmospheric pressure

(= factor 1.1)
Switching characteristic: High Trip Point
Hysteresis: 10 mbar

Measurement signal
(pressure p)

1078 mbar

980 mbar

Atmospheric pres/
sure (ambience)

ATM setpoint 110%
(= factor 1.1)

Hysteresis (1% F.S.)

Off

On

Off

Threshold value

Reset value

8

)

The atmospheric pressure sensor measures the atmospheric pressure

(pressure outside the vacuum system and can be calibrated against the
diaphragm capacitive sensor in the gauge (

→  Fehler! Textmarke

nicht definiert.

).