• the digital input "Remote Zero" (briefly apply the

supply voltage to pin 10),

• an INFICON Vacuum Gauge Controller (VGC series).

While the gauge is being heated and/or under atmos-

pheric pressure, the zeroing function is locked in order
for operating errors to be prevented.

Evacuate the gauge to a pressure according to the table

below:

F.S.

Recommended final pressure for

zero adjustment

1100 mbar
1000 Torr
500 Torr/mbar
200 Torr/mbar
100 Torr/mbar
50 Torr/mbar
20 Torr/mbar
10 Torr/mbar
5

Torr/mbar

2

Torr/mbar

1

Torr/mbar

0.5 Torr/mbar
0.25 Torr/mbar
0.1 Torr/mbar
0.05 Torr/mbar
0.02 Torr/mbar
0.01 Torr


<5×10

-2

Torr

<2.5×10

-2

Torr

<10

-2

Torr

<5×10

-3

Torr

<2.5×10

-3

Torr

<10

-3

Torr

<5×10

-4

Torr

<2.5×10

-4

Torr

<10

-4

Torr

<5×10

-5

Torr

<2.5×10

-5

Torr

<10

-5

Torr

<5×10

-6

Torr

<2.5×10

-6

Torr

<10

-6

Torr

<5×10

-7

Torr

<6.65×10

0

Pa

<6.65×10

0

Pa

<3.33×10

0

Pa

<1.33×10

-0

Pa

<6.65×10

-1

Pa

<3.33×10

-1

Pa

<1.33×10

-1

Pa

<6.65×10

-2

Pa

<3.33×10

-2

Pa

<1.33×10

-2

Pa

<6.65×10

-3

Pa

<3.33×10

-3

Pa

<1.33×10

-3

Pa

<6.65×10

-4

Pa

<3.33×10

-4

Pa

<1.33×10

-4

Pa

<6.65×10

-5

Pa

<6.65×10

-2

mbar


<3.33×10

-2

mbar

<1.33×10

-2

mbar

<6.65×10

-3

mbar

<3.33×10

-3

mbar

<1.33×<10

-3

mbar

<6.65×10

-4

mbar

<3.33×10

-4

mbar

<1.33×10

-4

mbar

<6.65×10

-5

mbar

<3.33×10

-5

mbar

<1.33×10

-5

mbar

<6.65×10

-6

mbar

<3.33×10

-6

mbar

<1.33×10

-6

mbar

If the final pressure is too high for zero adjustment (>25%
of the F.S.), the zero cannot be reached and the
<STATUS> LED blinks green. If this is the case, activate
the factory setting and adjust the zero again (

→  34).

Operate the gauge for at least 2 hours (until the signal is

stable).