4 error status – INFICON Stripe CDG045Dhs 0.01 ... 1000Torr / mbar, new 10mTorr User Manual

Page 32

Advertising
background image

32

tina84e1 (2014-07)

If the <ZERO> button is released for more than 5 s, the

gauge returns the measurement mode.

The factory setting of the reset value is 1% F.S. above

the Low Trip Point and 1% F.S. below the High Trip
Point (hysteresis).

If after programming of the hysteresis the button <SP> is

pushed, the factory setting of the hysteresis (1%) is
reactivated.

Programming setpoint SP1

Programmable parameters:
(

→  [3], [4])

Low Trip Point
Low Trip Enable
Low Trip Point Hysteresis

High

Trip

Point

High Trip Enable
High Trip Point Hysteresis

Setpoint

Mode

Adjusting setpoint SP2

The adjustment procedure is the same as for setpoint SP1.


4.4 Error

Status

The setpoints SP1 and SP2 can be programmed to error status
via the diagnostic port or the EtherCAT interface.