8 Technical data
92
8.7
Test voltages according to EN 60730, part 1
8.8
Electrical safety
8.9
Environmental influences
Input and output against voltage supply
- At a voltage supply AC 110 to 240 V +10 % /-15 %
3.7 kV/50 Hz
- At a voltage supply AC/DC 20 to 30 V, 48 to 63 Hz
3.7kV/50 Hz
Clearances / creep paths
Mains to electronic components
and probes
≥
6 mm
/ ≥
8 mm
Mains to relays
≥
6 mm
/ ≥
8 mm
Relays to electronic components
and probes
≥
6 mm
/ ≥
8 mm
Electrical safety
According to DIN EN 14597 (DIN EN 60730-2-9)
Overvoltage category III, pollution degree 2
Protection rating I
With internal separation from SELV current circuits
Ambient temperature range
0 to. +55 °C
Storage temperature range
-30 to +70 °C
Temperature influence
≤
±
0.005 % / K dev. from 23 °C
1
for RTD temperature probes
≤
± 0.01% / K dev. from 23°C
1
for thermocouples and current
Environmental performance
85 % rel. humidity without condensation
(3K3 with extended temperature range according to DIN EN 60721-3-3)