7. Daily Use of the Instrument

7.1 Specimen freezing

• Select the sectioning temperature (chamber temperature) according to

the tissue type to be sectioned (

see "7.4 Temperature selection chart"

).

7.1.1 Quick freeze shelf

The cryochamber is equipped with a quick freeze shelf (5,

Fig. 23

) on which

up to 10 specimen stages with specimens can be attached.

The temperature of the quick freeze shelf is always lower than the cryo-

chamber temperature.

• Cut the specimen to size.

• Apply a sufficient amount of cryocompound

to a specimen disc to orient and mount the

specimen.

• Place the specimen on the disc and orient.

• Place the specimen disc in one of the holes of

the quick freeze shelf and freeze the specimen

at a low temperature.

• Once the specimen is frozen, insert the speci-

men disc (

Fig. 24

) with the specimen into the

specimen head (

Fig. 24

) and start sectioning.