Troubleshooting, 1 error messages in the display, Error messages in the display – Leica Biosystems CM1520 User Manual

Page 51

Advertising
background image

51

Leica CM1520 – Cryostat

If errors occur, they are output as error codes in the real-time display during normal operation

with the format EO:XX.

8. Troubleshooting

8.1 Error messages in the display

Error messages are displayed on the clock panel as follows EO: XX . The

following error messages might occur during operation:

Fig. 50

• In normal operation, the error displayed can be acknowledged by pressing any key once. Error

codes 21-28 are self-acknowledging if their cause has been eliminated.

• After 15,960 hours, "HELP" is output in the real-time display to indicate the need for service mainte-

nance. The display alternates between "HELP" and the real-time display. The note is acknowledged

by pressing a key on the keyboard. The "HELP" display reappears for a longer period.

• After 17,610 hours, "HELP" is output in the real-time display to indicate the need for service mainte-

nance. The display alternates between "HELP" and the real-time display. The note is acknowledged

by pressing a key on the keyboard. The "HELP" display reappears for a shorter period.

Error

Cause

Remedy

20

Calibration error; possibly defective controller board.

Switch the instrument on again.

If the error is displayed again:

Call service.

21

Clock battery on the controller board empty.

Call service.

23

Cryochamber temperature out of display range of

35 °C to -55 °C.

Eliminate cause.

25

Temperature sensor for cryochamber temperature defective. Call service.

27

Temperature sensor for defrost limiter defective.

Call service.

28

Both light barriers in the handwheel active.

Call service.

29

Service interval warning after 15960 hours.

Call service.

30

Service interval expired after 17610 hours.

Call service.

Advertising