Daily use of the instrument – Leica Biosystems CM1520 User Manual

Page 38

Advertising
background image

38

Instructions for Use V1.3 – 10/2012

7.

Daily Use of the Instrument

You can adjust the height of the anti-roll guide system using the knurled nut (8):
• If you turn the nut counterclockwise, the anti-roll guide system moves toward the blade.

• If you turn the nut clockwise, the anti-roll guide system moves away from the blade.
If the anti-roll guide system is in the wrong position relative to the blade, the following problems

will result:

Fig. I: The section rolls over the glass

insert of the anti-roll guide system.
Error: Glass insert not high enough.
Remedy: Turn the knurled nut coun-

terclockwise until the section is

pushed between the blade and anti-

roll guide as shown in Fig. III.

Fig. II: Section tears and block hits

the glass insert after sectioning.
Error: Anti-roll guide system is set

too high.
Remedy: Turn the knurled nut clock-

wise until the section is pushed be-

tween the blade and anti-roll guide

as shown in Fig. III.

Fig. 36,

Blade holder CE

I

II

III

Adjusting the anti-roll guide system

8

We ALWAYS recommend pre-adjusting the anti-roll guide system at a high section thick-

ness (e.g. 10 µm). – Start from there and work your way down to the desired section thick-

ness in small increments, readjusting the anti-roll guide system at each increment using

the knurled nut.

Advertising