Daily use of the instrument – Leica Biosystems CM1520 User Manual

Page 45

Advertising
background image

45

Leica CM1520 – Cryostat

Blade holder CE

1. To release the segment arc (2) from the base,

use a 4 mm Allen key to loosen the screw

over the clearance angle adjustment (23) and

remove the segment arc (2) from the blade/

knife holder base.

2. Fold the anti-roll guide system (4) to the left

while holding the lever (11).

3. Open the clamping lever (10,

Fig. 47

) by turning

it counterclockwise and pull it out.

4. Afterwards, the pressure plate (36) can be

removed for cleaning (with alcohol).

5. Open the clamping lever of the blade holder

(12,

Fig. 47

) by turning it counterclockwise and

pull it out. The blade holder can now be moved

sideways and can be taken out of the segment

arc.

36

4

11

Fig. 47

12

Knife holder CN

Fig. 48

Occasionally oil the moving parts such as the

axes (37) and (38) and the slot (39) with a drop of

cryostat oil (Type 407).

39

38

37

Disinfection

Spray contaminated surfaces with a uniform coat

of concentrate or wipe them down with a rag

soaked in it. After the recommended exposure

time, wipe it off carefully.

7. Daily Use of the Instrument

10

23

2

Place the removed parts on an absorbent cloth in

the drying chamber (up to max. 65 °C) and let the

paraffin contamination run off.

If several blade/knife holders are

cleaned at the same time, the parts

must NOT be mixed up! Failure to ad-

here to this may result in sectioning

problems!

There is a burn hazard when remov-

ing the parts from the drying chamber

(65 °C). Wearing safety gloves is recom-

mended!

Also observe the warning message on

Page 13!