Leica CM1520 – Cryostat

57

8. Troubleshooting

The battery is exchanged during routine maintenance.
If the customer has decided against a maintenance contract, the battery must be replaced after

7 years at the latest. All settings (time, defrost time, section thickness etc.) are lost when the

instrument is switched off.
There is no danger for the instrument. However, the customer must re-enter the setting values

after changing the battery.

8.3.1 Changing the battery

Problem

Cause

Remedy

No or insufficient refrigeration

- Compressor defective.

- Leak in the cooling system.

- Inappropriate site conditions.

- Ventilation grilles of the condens-

er dirty.

- Call technical service.

- Call technical service.

- Check Installation site re-

quirements (item 4.1)

- Clean air inlet opening.

Scraping noise at the slot cover of

the microtome

- Friction between the slot cover

and the microtome housing.

- Apply cryostat oil to the slit

cover and distribute it by

moving the handwheel or

using a clean cloth.