Daily use of the instrument – Leica Biosystems CM1520 User Manual

Page 41

Advertising
background image

41

Leica CM1520 – Cryostat

7. Daily Use of the Instrument

Inserting/removing the knife

24

3

3a

3a

7

7

• Insert the long knife support (3) on the side

above the knurled screw (3a) so that the

recess (24) faces the user – turn the height

adjustment knurled screws until the lower stop

is reached.

• The knife can now be inserted from the side

and its height adjusted via the knurled screws

(3a). The upper edge of the rear clamping jaw

serves as an index point for the correct knife

height. The height of the knife blade should

match that of the rear clamping jaws. This

makes it possible to accurately adjust even

greatly worn knives up to a height of 25 mm.

• Once the correct height is obtained, tighten

the wingscrews (7) in alternating order until

both are tight.

• To remove the knife, follow these instructions

in opposite order.

Open the wingscrews by turning them coun-

terclockwise and then pull the knife out to the

side.

Fig. 41

Fig. 42

After removing the knife from the knife

holder, place it safely into the knife

case. NEVER place it onto the work

surface next to the instrument!

The height of resharpened knives must be adjusted using knurled screws (3a) (approx.

1mm under the edge of the clamping jaw).

Make sure that the knife is parallel from one end to the other.

The safety gloves included in the standard scope of delivery must be worn when insert-

ing/ejecting the knife!

Advertising