7.

Daily Use of the Instrument

Fig. 43

The knife guard (18) is fixed and integrated into the

clamping jaws. The knife guard has handles (19)

that allow it to be moved. The knife guard is suit-

able for knife lengths up to 16cm. Always cover

exposed parts of the knife blade after sectioning.
The anti-roll guide system can be moved sideways

(only for the 84 mm variant). To better find the mid

position, a groove (17) is provided in the shaft (16),

in which the anti-roll guide sits.

19

19

18

16

17

Knife guard/lateral movement for knife holder CN

Clearance angle adjustment

The clearance angle adjustment for the knife holder CN is carried out exactly as described for

the blade holder CE.

Adjusting the anti-roll guide

The anti-roll guide adjustment for the knife holder CN is carried out exactly as described for the blade

holder CE. For lateral adjustments, refer to the section on lateral movement (on this page above).

• Pressure plate permits the full length of the

knife to be utilized.

• Use carbide metal and steel knives here.