Data Acquisition

Validate

120

Data acquisition with static load

To acquire data at static load levels:

Note

Before you start, enter a Sample Points value sufficient to allow data
sampling at the various static levels. A value of 100 should suffice in
most cases.

Note

Taking readings near zero should be done only when using bending
strain validation criteria. Avoid using readings near zero if percent level is
used.

1. Start at 0 load. Press the Start button then press the Stop button after several

samples of data have been taken.

2. Manually ramp to (nominally) +50% full scale. Press the Resume button

then press the Stop button after several samples of data have been taken.

3. Manually ramp to +100% full scale. Press the Resume button then press the

Stop button after several samples of data have been taken.

4. Manually ramp to (nominally) +50% full scale. Press the Resume button

then press the Stop button after several samples of data have been taken.

5. Manually ramp to 0 load. Press the Resume button then press the Stop

button after several samples of data have been taken.

6. Manually ramp to (nominally) -50% full scale. Press the Resume button

then press the Stop button after several samples of data have been taken.

7. Manually ramp to -100% full scale. Press the Resume button then press the

Stop button after several samples of data have been taken.

8. Manually ramp to (nominally) -50% full scale. Press the Resume button

then press the Stop button after several samples of data have been taken.

9. Manually ramp to 0 load. Press the Resume button then press the Stop

button after several samples of data have been taken.

Repeat for a second cycle if desired.

= take data

0

+100%

full scale

-100%

full scale

*

*

*

* Taking readings near zero
should be done only when
using bending strain validation
criteria. Avoid using readings
near zero if percent level is
used.