MTS 709 Alignment System User Manual

Page 91

Advertising
background image

Get Specimen Offset Values

709 Alignment System

Specimen Offsets

91

5. Check the bent mean offset values.

The bent mean values are the average of both the bent front and bent
back values taken in steps 3 and 4. These values will be used when the
Specimen Bent selection is active in the Use Offset section of the
Align tab.

Also included are the X and Y bending components due to the specimen
and the load unit:

bending strain due to straightening the specimen (bent mean offset
relative to the gauge offset)

Xs - bending strain in the x-axis
Ys - bending strain in the y-axis

bending strain due to load unit misalignment (bent front offset relative
to the bent mean offset)

Xm - bending strain in the x-axis
Ym - bending strain in the y-axis