Alignment procedure overview, Alignment procedure overview 21 – MTS 709 Alignment System User Manual

Page 21

Advertising
background image

709 Alignment System

Alignment Procedure Overview

21

Alignment Procedure Overview

Following is an overview of a typical alignment procedure. It is intended as a
guide and will apply to most system configurations.

1. Check the controller setup and tuning before inserting the alignment

specimen instrumented with strain gauges: With no specimen clamped in the
grips, verify the stroke control tracks without drift. With an expendable
specimen clamped in the grips, verify the load control smoothly tracks a
slow (less than 1 Hz) function generator ramp.

2. Insure that the specimen is properly connected to the 709 electronics. Wait

20 to 30 minutes to insure there are no temperature gradients.

3. Confirm the specimen geometry and end levels in the specimen tab. The end

levels are typically set to the Load @ 2000 µe shown in the second column.

4. In the align tab, check the specimen offsets. Do this by toggling between

None, Current Tare, and Specimen Gauge. Current Tare and Specimen
Gauge
(stored and averaged offsets) should be very close (within 5%). If
they are not, the Specimen Gauge offsets should be cleared and re-
established.

5. With the system in displacement control install the specimen in the lower

grip (or upper grip if desired). Clamp that end of the specimen in the grip.

6. Zero the load cell.

7. Move the actuator until the other end of the specimen is in the proper

location of the other grip is in position to clamp.

8. Now check to see if the Bent Offset is properly established. Do this by

toggling between Current Tare and Bent Offset. Normally, you should see
very little change between Current Tare and Bent Offset. Note that Bent
Offset
is not really needed unless the specimen is not straight (i.e. has
manufacturing eccentricities or has been deformed).

Advertising