6.1 Problems and their solutions

■■■■■■■■■■■■■■■■■■■■■■

92

■■■■■■■■

882 Compact IC plus – Anion – MCS

Problem

Cause

Remedy

MCS – not connected.

Connect the MCS.

Peristaltic pump –
insufficient or no
delivery rate

Peristaltic pump – contact
pressure too weak.

Correctly set contact pressure (see "Set flow
rate", page 44)
.

Peristaltic pump – filter
blocked.

Replace the filter (see Chapter 5.10.2.2, page
79)
.

Peristaltic pump – pump
tubing defective.

Replace pump tubing (see Chapter 5.10.2.1,
page 78)
.

Very noisy baseline

High pressure pump – con-
taminated pump valves.

Clean pump valves (see Chapter 5.5.2, page
64)
.

Eluent – Leak in the eluent
path.

Check the eluent path.

Eluent – Blockage in the
eluent path.

Check the eluent path.

High pressure pump –
defective piston seals.

Replace (see Chapter 5.5.2, page 64) piston
seals .

MCS –
CO

2

adsorption cartridge

exhausted.

Replace CO

2

adsorption cartridge (see Chapter

5.12.1, page 87).

Pulsation absorber not
connected.

Connect pulsation absorber (see Chapter 3.11,
page 35)
.

Pulsation absorber not
connected. or defective.

Connect pulsation absorber (see Chapter 3.11,
page 35)
. or replace it.

MCS – vacuum pump
defective.

Contact the Metrohm Service.

Background conduc-
tivity too high

Suppressor – not connec-
ted.

Connect the suppressor (see Chapter 3.14.1,
page 45)
.

MCS – not connected.

Connect the MCS.

Incorrect eluent.

Change the eluent (see Chapter 5.4.2.3, Page
63)
.

Suppressor – regeneration
or rinsing solution flow
problems.

Check the flow of the regeneration solution
and the rinsing solution.