3.2 Initial installation

■■■■■■■■■■■■■■■■■■■■■■

12

■■■■■■■■

882 Compact IC plus – Anion – MCS

Prepare a pump tubing for the rinsing solution:

– Plug a tubing olive (6.2744.034) onto one end of the pump

tubing (6.1826.320).

– Plug a pump tubing connection (6.2744.180) onto the

other end of the pump tubing.

– Connect one end of the aspirating capillary (6.1803.020)

for the rinsing solution to the tubing olive on the pump
tubing using a short pressure screw (6.2744.070).

– Guide the other end of the aspirating capillary out of the

instrument through a capillary feed-through, slide it
through a bottle attachment (6.1602.150) and screw the
bottle attachment onto the bottle (6.1608.020) containing
the rinsing solution. Ensure that the end of the aspirating
capillary reaches down to the bottom of the bottle.

– Place the pump tubing into the other tubing cartridge.

Place both tubing cartridges into the peristaltic pump.

6 Connecting the MSM

(see Chapter 3.14, page 45)

Connect the capillary of the MSM labeled with regenerant to the
pump tubing connection of the pump tubing for the regeneration
solution using a short pressure screw (6.2744.070).

Connect the capillary of the MSM labeled with rinsing solution
to the pump tubing connection of the pump tubing for the rinsing
solution using a short pressure screw (6.2744.070).

Guide the two capillaries of the MSM labeled with waste reg.
and waste rins. out of the instrument through a capillary feed
through to a waste container and fasten them there.

7 Connecting the MCS

(see Chapter 3.15, page 48)

Attach the CO

2

adsorption cartridge (6.2837.000) to the adsorp-

tion cartridge holder (6.2057.080) (see "Installing the adsorption
cartridges", page 51)
.

Prepare the H

2

O adsorption cartridge (6.2837.010) (see leaflet to

the H

2

O adsorption cartridge) and attach it to the adsorption car-

tridge holder as well (see Figure 26, page 50).

Plug the adapter (6.1808.190) onto the PVC tubing and connect
the two adsorption cartridges with one another (see Figure 26,
page 50)
.

Place the adsorption cartridge holder (6.2057.080) in the detector
chamber.

Connect the MCS air aspirating capillary to the tip of the CO

2

adsorption cartridge (6.2837.000).