3 Installation

882 Compact IC plus – Anion – MCS

■■■■■■■■

13

8 Connecting the instrument

■

Connect the instrument to a computer on which the software
MagIC Net™ is installed using the USB cable (6.2151.020) (see
Chapter 3.16.1, page 52)
.

■

Connect the instrument to the mains supply (see Chapter 3.16.2,
page 52)
.

9 Initial start-up

(see Chapter 4.1, page 57)

■

Switch on the PC and start the software MagIC Net™.

■

Switch on the instrument.

■

Deaerate the high pressure pump (see Chapter 3.9.2, page 32).

■

Set contact pressure of the peristaltic pump (see "Set flow rate",
page 44)
.

■

Rinse the instrument without column with eluent for 5 minutes.

10 Installing guard and separation column

■

Remove the coupling (6.2744.040) between the column inlet
capillary and the capillary of the MSM labeled with in.

■

Optional Connect guard column (see Chapter 3.17, page 53)

– Fasten the guard column to the end of the column inlet

capillary (see leaflet to the guard column).

– Rinse the guard column with eluent for approx. 5 minutes.

■

Connect the separation column (see Chapter 3.18, page 54)

– Connect the inlet oft the separation column either with the

end of the column input capillary using a short pressures
screw (6.2744.070).
OR
Connect the inlet of the separation column with the guard
column (if used) (see leaflets to the separation column and
the guard column)

– Connect the MSM capillary labeled with in with the output

of the separation column using a short pressure screw
(6.2744.070).

■

Hang separation column with chip in the column holder of the
instrument.

11 Conditioning the instrument

(see Chapter 4.2, page 58)

The instrument is now ready for measuring samples.