3.8 Eluent
■■■■■■■■■■■■■■■■■■■■■■
28
■■■■■■■■
882 Compact IC plus – Anion – MCS
■
Fasten the completely equipped bottle attachment (10-2) on the
eluent bottle (12-10). The aspiration filter (12-6) must rest on the
base of the eluent bottle.
■
Close the remaining small opening on the bottle attachment with
a threaded stopper (12-14) from the accessories set.
6 Mounting the adsorber tube
Note
If alkaline eluents and eluents with lower buffer capacity are used,
the eluent bottle must be equipped with an adsorber tube filled
with CO
2
■
First, place a piece of cotton (12-3), then the CO
2
in the large opening of the adsorber tube (12-2) and close with
the plastic cover.
■
Fasten the adsorber tube (12-2) onto the bottle attachment
(12-11) using the SGJ clip (12-12).
1
1
2
3
4
5
6
7
8
9
10
11
12
13
14
Figure 12
Eluent bottle – connected
2
Eluent aspiration tubing (6.1834.080)
For aspirating the eluent. Pre-installed.
2
Adsorber tube (6.1609.000)