4.2 Conditioning

■■■■■■■■■■■■■■■■■■■■■■

58

■■■■■■■■

882 Compact IC plus – Anion – MCS

4 Deaerate the high pressure pump

■

Deaerate the high pressure pump(s) via the purge valve (see
Chapter 3.9.2, page 32)
.

5 Set the contact pressure of the peristaltic pump

Note

This work step needs to be performed only if a peristaltic pump is
being used.

■

If peristaltic pumps are used, set the contact pressure (see "Set
flow rate", page 44)
.

6 Rinsing the instrument without columns

■

Rinse the instrument (without columns) with eluent for 5 minutes.

The instrument is now ready for the installation of the columns (see Chap-
ter 3.17, page 53)
.

4.2

Conditioning

After the installation and after switching on the instrument, the system
must be conditioned with eluent until a stable baseline is reached.

Note

After a change of eluent (see Chapter 5.4.2.3, page 63), the condi-
tioning time can lengthen considerably.

Conditioning the system

1 Preparing the software

Caution

Ensure that the flow set is not higher than the flow permissible for
the corresponding column (see column leaflet and chip data set).

■

Start the PC program MagIC Net™.

■

Open the Equilibration tab in MagIC Net™.