3 Installation

882 Compact IC plus – Anion – MCS

■■■■■■■■

31

Note

The eluent aspiration capillary is already installed in the newly delivered
instrument. The following installation instructions need not be carried
out at the time of initial installation.

Connecting inlet to the high pressure pump

5

1

2

3

4

Figure 14

High pressure pump – Connect inlet

1

Pressure screw
For connecting the coupling (14-

2

) to the

pump head inlet capillary (13-

7

).

Can be ordered together with the coupling
under the number (6.2744.230).

2

Coupling (6.2744.230)
For connecting the eluent connection capil-
lary (14-

4

) to the input of the high pressure

pump.

3

Clamping screw

4

Eluent aspiration tubing
Eluent aspiration tubing (6.1834.080) or
(6.1834.090).

5

Backup ring

1 Connecting coupling

Fasten the coupling (14-2) with a pressure screw (14-1) on the pump
head inlet capillary (13-7).

2 Connecting eluent aspiration tubing

Caution

The clamping screws must be tightened carefully. To tighten, grip
the coupling (14-2) with the key (6.2739.000) and grip the clamp-
ing screw (14-3) with the wrench (6.2621.050).

Plug the eluent aspiration tubing (14-4) into the coupling (14-2).

Tighten the clamping screw (14-3).