5 Operation and maintenance

882 Compact IC plus – Anion – MCS

■■■■■■■■

75

Changing the filter

The flow must be stopped before changing the filter.

1 Removing the inline filter

Unscrew the pressure screws (36-1) from the inline filter.

2 Unscrewing the filter screw

Screw the filter screw (36-4) out of the filter housing (36-2) with
the aid of two adjustable wrenches (6.2621.000).

3 Inserting the filter

Remove the old filter (36-3) using tweezers.

Place the new filter (36-3) flat in the filter housing (36-2) using
tweezers.

4 Mounting filter screw

Screw the filter screw (36-4) back into the filter housing (36-2)
and tighten by hand. Then additionally tighten slightly with two
adjustable wrenches (6.2621.000).

5 Remounting the inline filter

Screw pressure screws (36-1) back onto the inline filter.

6 Rinsing the inline filter

Dismantle the guard column (if present) and the separation col-
umn and replace with a coupling (6.2744.040).

Rinse the instrument with eluent.