Metrohm 882 Compact IC plus – Anion – MCS User Manual

Page 21

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

882 Compact IC plus – Anion – MCS

■■■■■■■■

11

Connect the leak sensor (see Chapter 3.6.2, page 20).

Connect the drainage tubings (see Chapter 3.6.3, page 21).

3 Connecting the eluent path

Lead the eluent aspiration tubing (6.1834.080) out of the instru-
ment through a capillary feed-through and connect it with the
eluent bottle (see Chapter 3.8, page 25).

Connect the column inlet capillary (6.1831.150) and the capillary
of the MSM labeled with in to one another using a coupling
(6.2744.040) and two short pressure screws (6.2744.070).

Use a long pressure screw (6.2744.090) to connect the capillary
of the MSM labeled with out to the input of the of the MCS (see
"Connecting the MCS", page 49)
.

Connect the detector inlet capillary with a long pressure screw
(6.2744.090) to the output of the MCS (see "Connecting the
MCS", page 49)
.

4 Connecting the sample path

Guide the sample aspiration capillary connected to the sample
input of the injection valve out of the instrument through a capil-
lary feed-through and connect it with the Sample Processor, if
applicable (see Sample Processor manual).

Guide the sample outlet capillary connected to the sample output
of the injection valve out of the instrument through a capillary
feed-through and onward to the waste container and then fasten
it there.

5 Installing the peristaltic pump

(see Chapter 3.13.2, page 41)

Prepare the pump tubing for the regeneration solution:

– Plug a tubing olive (6.2744.034) onto one end of the pump

tubing (6.1826.320).

– Plug a pump tubing connection (6.2744.180) onto the

other end of the pump tubing.

– Connect one end of the aspirating capillary (6.1803.020)

for the regeneration solution to the tubing olive on the
pump tubing using a short pressure screw (6.2744.070).

– Guide the other end of the aspirating capillary out of the

instrument through a capillary feed-through, slide it
through a bottle attachment (6.1602.150) and screw the
bottle attachment onto the bottle (6.1608.020) containing
the regeneration solution. Ensure that the end of the aspi-
rating capillary reaches down to the bottom of the bottle.

– Place the pump tubing into a tubing cartridge.