Figure 27, Pump head – removing the piston – Metrohm 882 Compact IC plus – Anion – MCS User Manual

Page 75

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Operation and maintenance

882 Compact IC plus – Anion – MCS

■■■■■■■■

65

Cleaning/replacing the zirconium oxide piston

Clean one piston after the other as follows:

1 Removing the piston cartridge from the pump head

Loosen the piston cartridge with a wrench and unscrew from the
pump head by hand.

2

1

Figure 27

Pump head – removing the piston

1

Pump head

2

Piston cartridge

2 Dismantling the piston

Caution

On the inside of the piston cartridge there is a taut spring than can
jump out of the piston cartridge if suddenly loosing tension.

When opening the piston cartridge, hold pressure towards the
spring and unscrew carefully.

■

Loosen the screw of the piston cartridge with a wrench and
unscrew carefully by hand and by holding pressure towards the
taut spring.

■

Remove the zirconium oxide piston and lay on a tissue.

■

Remove the spring retainer, spring and the inner plastic sleeve
from the piston cartridge and lay by.

Advertising