3.18 Separation column

■■■■■■■■■■■■■■■■■■■■■■

56

■■■■■■■■

882 Compact IC plus – Anion – MCS

2 Rinsing the separation column

■

Place beaker under the outlet end of the separation column.

■

Set the flow rate of the high pressure pump according to the data
given in the leaflet of the separation column.

■

Start the high pressure pump and rinse the separation column
approx. 10 minutes with eluent.

■

Switch off the high pressure pump again.

3 Mounting the separation column

■

Fasten the column output capillary to the output of the separation
column using a PEEK pressure screw (6.2744.070).

■

Hang separation column with chip into the column holder.

Note

The iColumns are equipped with a chip on which their operating data is
saved. The chip has to be hooked into the chip holder provided for this
so that the column recognition can function.