4.4 Methods
761 Compact IC
115
Set baseline point
without Set baseline point
at 4.5, 6, 7.2, 8.9
and 9.8 min
Set baseline point
Set defined baseline point for better evaluation
of peaks on descending or ascending base-
lines. The baseline between two successive
baseline points is set to zero for optimum inte-
gration.
Force horizontal
Cancel horizontal
baseline
baseline
Force horizontal baseline
Set horizontal baseline for a single peak. The
baseline is drawn forwards from the peak start
point. The intersection of the baseline with the
signal is defined as peak end point.
Cancel horizontal baseline
Clear
Force horizontal baseline
mode.
Force horizontal
Cancel horizontal
base back
base back