Metrohm 761 Compact IC User Manual
Page 42

2 Installation
761 Compact IC
32
74
73
75 76 77 76 75 78 75
54
28
79 80
81
Fig. 12: Installation of precolumn cartridges with cartridge head
28 Column connection capillary from
injector
77
Precolumn cartridge
54
Compression fitting (6.2744.010)
78
Steel spacer (6.2821.080)
73
Manufit housing
79
4 Steel meshes (6.2821.020)
74
Steel connector
(of IC separating column)
80 Manufit pressure screw
75 PTFE gasket (6.2821.010)
81 IC separating column
76
2 Steel meshes (6.2821.020)
3 Connect precolumn
•
Fit PEEK compression fitting 54 to the column connection
capillary 28 mounted on the injection valve (see section
2.3.5).
•
Screw column connection capillary 28 to Manufit housing 73.
The precolumn cartridge built into the cartridge head can only be
rinsed together with the separating column (see section 2.7.7/8). This
increases the rinsing time by approx. 20 min.