Metrohm 761 Compact IC User Manual

Page 42

Advertising
background image

2 Installation

761 Compact IC

32

74

73

75 76 77 76 75 78 75

54

28

79 80

81

Fig. 12: Installation of precolumn cartridges with cartridge head

28 Column connection capillary from

injector

77

Precolumn cartridge

54

Compression fitting (6.2744.010)

78

Steel spacer (6.2821.080)

73

Manufit housing

79

4 Steel meshes (6.2821.020)

74

Steel connector

(of IC separating column)

80 Manufit pressure screw

75 PTFE gasket (6.2821.010)

81 IC separating column

76

2 Steel meshes (6.2821.020)


3 Connect precolumn

•

Fit PEEK compression fitting 54 to the column connection

capillary 28 mounted on the injection valve (see section

2.3.5).

•

Screw column connection capillary 28 to Manufit housing 73.


The precolumn cartridge built into the cartridge head can only be

rinsed together with the separating column (see section 2.7.7/8). This

increases the rinsing time by approx. 20 min.