3.3 Calibration
761 Compact IC
57
9 Enter information for determination
⇒
Click on
or select
Passport
of the
Method
menu. The
passport opens with the
General
page, which contains the
general information about the determination.
⇒
Enter a title for the calibration as
Ident
(e.g.
Standard
).
⇒
Change the analysis time under
Duration
if required.
⇒
Change other parameters on the passport tabs to character-
ize sample and acquirement conditions if required.
10 Switch injection valve to "FILL" position
⇒
Click on the
button of the 761 icon in the system
window to switch the injection valve to the "FILL" position. At
the same time, the suppressor module is switched to the next
position.