2 Installation

761 Compact IC

48

6 Condition system

•

Rinse the system with eluent until the desired stability of the

baseline is reached (normally 30

…

60 min; if the eluent is

changed, the establishment of the ion exchanger equilibrium

on the separating column can take longer). After this time the

suppressor module is sufficiently conditioned too.

•

The instrument is now ready for sample determinations using

the selected system.

Pump tubings are consumable material with a lifetime which depends

on the contact pressure. This is why the tubing cartridges should be

raised completely by loosening snap-action lever on the right-hand

side if the pump is to remain switched off for a considerable length of

time (the set contact pressure remains unchanged).