2 Installation
761 Compact IC
44
5 Inlet capillary for H
2
O
•
Attach inlet capillary 98 marked with "H
2
O" of suppressor
module 47 using a 6.2744.010 Compression fitting to the filter
unit PEEK 36 connected to the front pump tubing 93.
6 Outlet capillary for H
2
O
•
Pull outlet capillary 100 marked with "Waste" of the suppres-
sor module 47 from below through one of the openings 13
out of the inner compartment of the 761 Compact IC.
•
Lead outlet capillary 100 to a sufficiently large waste con-
tainer and fix it in place.
7 Fasten capillaries to the side walls
•
If necessary, the two aspirating tubings 89 and 90 can be
fixed in the required position in the interior with the help of a
Y.107.0150 self-adhesive strap.
•
If necessary, the two outlet capillaries 100 and 101 can be
fixed in the required position in the interior with the help of a
Y.107.0150 self-adhesive strap.