9 putting into operation, 1 putting into operation without suppressor – Metrohm 761 Compact IC User Manual
Page 55

2.9 Putting into operation
761 Compact IC
45
2.9
Putting into operation
2.9.1
Putting into operation without suppressor
Before sample solutions can be injected at the 2.761.0010 Compact
IC (without suppressor), the entire system must be tested for leaks and
then conditioned with eluent until the baseline is stable. Proceed as fol-
lows:
1 Open and connect system
•
Start the «761 Compact IC» PC program, if it has not already
been started (see section 2.5.3).
•
Select
File / Open / System
in the main window. Select the
system file suited for the inserted separating column (e.g.
n-
01.smt
) and click on
<Open>
.
•
Select the
Connect to workplace
item from the
Control
menu of
the system window.
2 Open control window
•
Double-click the 761 icon in the system window. The control
window for manual control of the 761 Compact IC appears,
which indicates conductivity, pressure and current system
parameters.