4 Operation
761 Compact IC
116
Force horizontal base back
Set horizontal baseline for a single peak. The
baseline is drawn backwards from the peak
end point. The intersection of the baseline with
the signal is defined as peak start point.
Cancel horizontal base back
Clear
Force horizontal base back
mode.
Enable baseline
Disable baseline
penetration
penetration
Enable baseline penetration
Allow crossing of the signal by the baseline.
Disable baseline penetration
Clear
Enable baseline penetration
mode. The
baseline is aligned to the bottom value of the
signal.
Integration events for integration parameters
Set width
Sets
new
Width
parameter that supersedes de-
fault linear growth of expected peak.
Set slope
Sets
new
Slope
value.
Set min height
Sets
new
Minimum peak height
value.
Set rider ratio
Sets new value of
Rider ratio
parameter.