Metrohm 761 Compact IC User Manual

Page 49

Advertising
background image

2.8 Suppressor module

761 Compact IC

39

•

Press contact pressure lever 49 on both tubing cartridges

down as far as it will go.

•

Insert the pump tubings 92 and 93 (6.1826.060) into each of

the tubing cartridges as shown in Fig. 15. The white-yellow

stopper 94 must click into the corresponding holder on the

left-hand side of the tubing cartridge.

•

Place the tubing cartridges on mounting pin 53 and press

down on the right-hand side until snap-action lever 51 clicks

into position on holding clamp 50. Take care that no kinks are

formed in the pump tubing.

2 Install Filter units PEEK

•

Mount a coupling 91 (6.2744.030) to the outlet end of the two

Pump tubings 92 and 93.

•

Attach a piece of PTFE tubing 95 (6.1803.020) cut to the

required length (normally ca. 10 cm) using a compression

fitting 54 (6.2744.010) to the other end of this coupling.

•

Attach the PTFE tubing 95 using a compression fitting 54

(6.2744.010) to the filter-screw 57 of the filter unit PEEK (see

section 2.3.6).

89/90

91 92/93 94

48

49

94 91

98/99

51

54

54

95 54

59

57 54

Fig. 15: Installing pump tubings

48

Tubing cartridge

91 Coupling (6.2744.030)

49

Contact pressure lever

92 Pump tubing (6.1826.060) for H

2

SO

4

51 Snap-action lever

93 Pump tubing (6.1826.060) for H

2

O

54

PEEK compression fitting

(6.2744.010)

94 Stopper (white-yellow)

57 Filter-Screw of Filter Unit

Part of 6.2821.120 Filter unit

95

PTFE tubing (6.1803.020)

59 Filter-Housing of Filter Unit

Part of 6.2821.120 Filter unit

98 Suppressor inlet capillary for H

2

O

89 Aspirating tubing for H

2

O

99 Suppressor inlet capillary for H

2

SO

4

90 Aspirating tubing for H

2

SO

4

Advertising