Metrohm 761 Compact IC User Manual
Page 97

4.3 Systems
761 Compact IC
87
mouse button and selecting the menu item
Choose color / ...
.
Actual values
Display of actual values
Flow, mL/min
Display of flow rate of the high-pressure pump.
Pressure max, MPa
Display of maximum pressure limit for high-
pressure pump.
Pressure min, MPa
Display of minimum pressure limit for high-
pressure pump.
Full scale,
µ
S/cm
Display of selected full scale range.
Remote lines
Display of current remote line settings.
Valve
Injection
valve
<Inject>
Switch injection valve to "INJECT" position.
<Fill>
Switch injection valve to "FILL" position
Suppressor
Suppressor
module
<Step>
Switch the suppressor module to the next posi-
tion. The time since the last switching of the
suppressor module is displayed in the field be-
side the
<Step>
button.
IC pump
High-pressure
pump
<On>
Start high-pressure pump.
<Off>
Stop high-pressure pump.
Peristaltic pump
Peristaltic
pump
<On>
Start peristaltic pump.
<Off>
Stop peristaltic pump.
System startup values
The system startup values are sent and applied
to the 761 Compact IC each time the system is
connected, a determination is started, or the
values are sent manually with
<Send to unit>
.
Flow, mL/min
Startup value for flow rate of the high-pressure
pump.
Entry range:
0.20
...
2.50 mL/min
Pressure max, MPa
Startup value for maximum pressure limit for
high-pressure pump. This limit is controlled
even without connection to the PC.
Entry range:
0.0
...
25.0 MPa
Pressure min, MPa
Startup value for minimum pressure limit for
high-pressure pump. This limit is controlled
even without connection to the PC.
Entry range:
0.0
...
25.0
MPa
Full scale,
µ
S/cm
Startup value for full scale range.
Selection:
50
,
250
,
1000
µ
S/cm
Remote lines
Startup value for remote line settings 1
…
8.
Selection:
0
,
1