5.2 Maintenance and servicing

761 Compact IC

181

•

Set the flow at the high-pressure pump to 0.5 mL/min and

rinse the suppressor unit with 1 mol/L H

2

SO

4

for 5 to 10 min.

•

As the pressure falls, slowly increase the flow at the high-

pressure pump to 2 mL/min. Do not exceed a maximum

pressure of 2 MPa (20 bar).

•

Switch off high-pressure pump.

•

Switch suppressor to the next position using the

button.

•

Connect the inlet capillary 98 marked "H

2

O" to the inlet capil-

lary 26 using coupling 33 (see Fig. 10 and Fig. 16).

•

Set the flow at the high-pressure pump to 0.5 mL/min and

rinse the suppressor unit with 1 mol/L H

2

SO

4

for 5 to 10 min.

•

As the pressure falls, slowly increase the flow at the high-

pressure pump to 2 mL/min. Do not exceed a maximum

pressure of 2 MPa (20 bar).

•

Switch off high-pressure pump.

3 Connect suppressor to IC system

•

Connect inlet capillaries 98 and 99 to the peristaltic pump

(see section 2.8.4).

•

If the pressure problems persist, replace the suppressor rotor

(see section 5.2.9).



5.2.8

Cleaning the suppressor

It may be necessary to clean the suppressor in the following cases:

•

High counterpressure on the suppressor connection

tubing

•

Irremediable blockage of the suppressor (the suppres-

sor can no longer deliver solutions)

•

Irremediable obstruction of the suppressor (the sup-

pressor can no longer be switched to next position)

To clean the connection piece and the rotor, proceed as follows (see

Fig. 22):

1 Disconnect suppressor from IC system

•

Disconnect input capillary 96 of the suppressor module 47

from the separating column 81 (see Fig. 16).

•

Disconnect output capillary 97 from inlet capillary 45.

•

Disconnect inlet capillaries 98 and 99 from the filter units 96

(supply from peristaltic pump).