2 Installation

761 Compact IC

28

3 Mount CO

2

absorber tube

•

First place a piece of cotton wool 70 followed by CO

2

ab-

sorber 69 (e.g. Merck soda-lime pellets with indicator, no.

6839.1000) in the large opening of absorber tube 72 and then

close it with the plastic lid.

•

Fasten absorber tube 72 to bottle attachment 66 with the aid

of SGJ clip 71.

4 Connect aspirating tubing to pump

•

Insert the free end of aspirating tubing 63 into one of the

openings 12 in the interior of the 761 Compact IC (see Fig. 2).

•

Pull aspirating tubing 63 sufficiently far into the interior of the

761 Compact IC, cut off to the required length and push at

least 5 mm of it onto aspirating capillary 40 (see Fig. 3 and

Fig. 4) of the high-pressure pump (it may be necessary to use

emery paper).

•

If necessary, fix aspirating tubing 63 in the required position

in the interior with the aid of the Y.107.0150 self-adhesive

strap.

Instead of threaded stopper

65

, the 4.420.0311 Tubing nipple (M6)

which belongs to the 6.1602.160 Bottle attachment accessories can

be used together with the included second E.301.0021 O-ring for

recycling operation (see section 5.2.3), in which the outlet capillary

from the detector block is led back into eluent bottle

67

.

63
64
65

66

72

68

71

70

69

67

Fig. 11: Connection of eluent bottle

63 Aspirating tubing (6.1834.010)
64 Tubing nipple (4.420.4300; M8)

with E.301.0021 O-ring

65 Threaded stopper (6.1446.040; M6)
66 Bottle attachment (6.1602.105)
67 Eluent bottle (6.1608.070)
68 Aspirating filter (6.2821.090)
69 CO

2

absorber

70 Cotton wool
71 SGJ clip (6.2023.020)
72 Absorber tube (6.1609.000)