5.2 Maintenance and servicing

761 Compact IC

175

2 Connect new column to injector

•

Remove end caps from column 81.

•

Screw inlet end of separating column 81 (note flow direction)

to column connection capillary 28 or to the precolumn (see

section 2.7.7/section 2.7.8).

3 Rinse column

•

Place beaker beneath the column outlet.

•

Switch on high-pressure pump and rinse column with eluent

for ca. 10 min, then switch off pump.

4 Connect column to detector block

•

Screw outlet end of separating column 81 to inlet capillary 45

or suppressor inlet capillary 96.

Changing the separation system
If you wish to replace an IC separating column by a column of a differ-

ent type, proceed as follows (see Fig. 14 and Fig. 16):

1 Remove old column

•

Switch off high-pressure pump and wait for pressure to drop.

•

Unscrew column 81 from inlet capillary 45 of the detector

block or from the suppressor inlet capillary 96.

•

Unscrew column 81 from column connection capillary 28 or

the precolumn.

2 Rinse with eluent

•

Place beaker beneath the column connection capillary 28.

•

Rinse IC system with eluent used for the separating column

(flow rate 1 mL/min) for ca. 15 min.

3 Connect new column to injector

•

Remove end caps from column 81.

•

Screw inlet end of separating column 81 (note flow direction)

to column connection capillary 28 or to the precolumn (see

section 2.7.7/section 2.7.8).

4 Rinse column

•

Place beaker beneath the column outlet.

•

Set flow rate for new separating column.

•

Switch on high-pressure pump and rinse column with eluent

for ca. 10 min, then switch off pump.

5 Connect column to detector block

•

Screw outlet end of separating column 81 to inlet capillary 45

or suppressor inlet capillary 96.