5.2.10 RS232 error rectification

Problem

Questions for corrective action

No characters can be re-
ceived on an attached
printer.

⇒

Are the devices switched on and the connecting cables
plugged in properly?

⇒

Is the printer set to ”on-line”?

⇒

Do the baud rate, data bits and parity for the two devices have
the same setting?

⇒

Is the handshake set properly?

If everything appears ok, try to print out a report by pressing the
<PRINT> key.

No data transmission oc-
curs and an error message
appears in the display of
the 766 IC Sample Proces-
sor.

⇒

⇒ RS error 36…

…39: Receive errors.

Do the RS232 data transmission parameters of the two de-
vices have the same setting?

⇒

RS error 40

…

…42: transmission errors.

Is the cable used correctly wired and plugged in?
Is the printer switched on and set to ”on-line”?

⇒

RS error 43: Data output of the 766 IC Sample Processor
blocked for longer than 3 s by XOFF.

The received characters
are garbled.

⇒

Do the data bits and parity of the two devices have the same
setting?

⇒

Does the baud rate of the two devices have the same setting?

⇒

Is the correct printer selected?

⇒

The data transfer has been interrupted by the hardware during
a printout. Set up connections again, switch printer off then
on.