Metrohm 766 IC Sample Processor User Manual

Page 130

Advertising
background image

6.5 Initialize data memory

766 IC Sample Processor

125

Prepare instrument for diagnostic test (see section 6.4.2).

If need be, press <9> key repeatedly until

diagnosis
>RAM initialization

Press <ENTER> key to open the following diagnostic menu:

>RAM initialization
select: param

ü

>RAM initialization
select: config

ü

>RAM initialization
select: setup

ü

>RAM initialization
select: assembly

ü

>RAM initialization
select: all

ь

ь

ь

ь

ь

Set method parameters with default values.

Ã

Set config parameters with default values.

Ã

Set setup parameters with default values.

Ã

Set assembly parameters with default values.

Ã

Deletes all user-defined methods

Ã

Pressing the <SELECT> key selects the submenus in turn. The in-

dividual initialization alternatives are accessed using the <ENTER>

key, exit is with the <QUIT> key.

The table shows which parts of the data memory are affected by the

corresponding initialization alternatives. When the instrument reacts

with a system crash after being switched on (undefined display, no

reactions to keystroke, etc.) we recommend to carry out the initiali-

zation with the submenu

'all'

.

If need be, press the <SELECT> key repeatedly until:

>RAM initialization
select: all

Press <ENTER>.

diagnosis
>RAM test

Press <QUIT>.

The instrument quits the diagnostic menu and runs a power on reset.

Advertising