6.5 Initialize data memory

766 IC Sample Processor

125

•

Prepare instrument for diagnostic test (see section 6.4.2).

•

If need be, press <9> key repeatedly until

diagnosis
>RAM initialization

•

Press <ENTER> key to open the following diagnostic menu:

>RAM initialization
select: param

ü

>RAM initialization
select: config

ü

>RAM initialization
select: setup

ü

>RAM initialization
select: assembly

ü

>RAM initialization
select: all

ь

ь

ь

ь

ь

Set method parameters with default values.

Ã

Set config parameters with default values.

Ã

Set setup parameters with default values.

Ã

Set assembly parameters with default values.

Ã

Deletes all user-defined methods

Ã

•

Pressing the <SELECT> key selects the submenus in turn. The in-

dividual initialization alternatives are accessed using the <ENTER>

key, exit is with the <QUIT> key.

•

The table shows which parts of the data memory are affected by the

corresponding initialization alternatives. When the instrument reacts

with a system crash after being switched on (undefined display, no

reactions to keystroke, etc.) we recommend to carry out the initiali-

zation with the submenu

'all'

.

•

If need be, press the <SELECT> key repeatedly until:

>RAM initialization
select: all

•

Press <ENTER>.

diagnosis
>RAM test

•

Press <QUIT>.

The instrument quits the diagnostic menu and runs a power on reset.