4.2 Basic settings

766 IC Sample Processor

53

>rack definitions 2
type: M129-2

M129-2

Type description of the rack
<SELECT> enables the choice of Metrohm-specific

and self-defined rack types (see section 4.5).

>rack definitions 2
work position 125 mm

0

125 mm

Working position of the needle
in mm from the upper stop.
Pressing <CLEAR> directly accepts the current lift po-

sition.

If you use the steel needle for 6.2743.050 sample tubes which are

sealed with 6.2743.060 PE caps, the working position must always be

set to 125 mm, since otherwise a vacuum can develop in the sample

tube and the sample will not be aspirated correctly.

>rack definitions 2
rinse position 125 mm

0

125 mm

Rinsing position of the needle
in mm from the upper stop.
Pressing <CLEAR> directly accepts the current lift po-

sition.

>rack definitions 2
shift position 0 mm

0 mm

Shifting position of the needle
in mm from the upper stop.
Only the value

'0'

is permissible.

>rack definitions 2
special position 0 mm

0

125 mm

Special position of the needle
in mm from the upper stop.
Pressing <CLEAR> directly accepts the current lift po-

sition.

>rack definitions 2
>>special positions

Submenu

Special positions submenu

>>special positions
special beaker 1 128

0

number of positions

Position of special beaker 1

>>special positions
special beaker 2 129

0

number of positions

Position of special beaker 2

etc. up to special beaker 8

Up to 8 special beaker positions can be defined

(see section 4.5).