6.4 Diagnosis
766 IC Sample Processor
121
Key code table
Code
Key
Code
Key
0
16
<7 / SAMPLE>
1
<HOLD / LEARN>
17
<4 / PUMP>
2
<STOP>
18
<1 / SCAN>
3
<START>
19
<0>
4
<CONFIG>
20
<END>
5
<PARAM>
21
<è>
6
<USER METHOD>
22
<CLEAR / RESET>
7
23
<ENTER>
8
<9 / LIFT>
24
<é>
9
<6>
25
<ê>
10
<3 / WAIT>
26
<SELECT / TOWER>
11
<* / ENDSEQ>
27
<QUIT>
12
<8 / MOVE>
28
<HOME>
13
<5>
29
<ç>
14
<2 / CTRL>
30
<INSERT >
15
<. / PRINT>
31
<DELETE>
6.4.6
Remote interface
This diagnosis step tests the functionality of all the output (14) and in-
put lines (8).
•
Prepare instrument for diagnosis (see section 6.4.2).
•
If necessary, press <ê> several times until
diagnosis
>remote test
•
<ENTER>
>remote test
remote test connector ?
•
Plug 3.496.8550 test plug into the remote interface without turning
the instrument off.
•
<ENTER>
The test runs automatically. If no error occurs, the following appears:
>remote test
remote test ok
•
Remove test plug and press <ENTER>.
diagnosis
>RS232 test