6.4 Diagnosis

766 IC Sample Processor

121

Key code table

Code

Key

Code

Key

0

16

<7 / SAMPLE>

1

<HOLD / LEARN>

17

<4 / PUMP>

2

<STOP>

18

<1 / SCAN>

3

<START>

19

<0>

4

<CONFIG>

20

<END>

5

<PARAM>

21

<è>

6

<USER METHOD>

22

<CLEAR / RESET>

7

23

<ENTER>

8

<9 / LIFT>

24

<é>

9

<6>

25

<ê>

10

<3 / WAIT>

26

<SELECT / TOWER>

11

<* / ENDSEQ>

27

<QUIT>

12

<8 / MOVE>

28

<HOME>

13

<5>

29

<ç>

14

<2 / CTRL>

30

<INSERT >

15

<. / PRINT>

31

<DELETE>

6.4.6

Remote interface

This diagnosis step tests the functionality of all the output (14) and in-

put lines (8).

Prepare instrument for diagnosis (see section 6.4.2).

If necessary, press <ê> several times until

diagnosis

>remote test

<ENTER>

>remote test
remote test connector ?

Plug 3.496.8550 test plug into the remote interface without turning

the instrument off.

<ENTER>

The test runs automatically. If no error occurs, the following appears:

>remote test
remote test ok

Remove test plug and press <ENTER>.

diagnosis
>RS232 test