3 Operating tutorial

766 IC Sample Processor

28

3.2

Configuration

CONFIG

ENTER

TOWER

SELECT

The dialog language can be set in the configuration menu. Press

<CONFIG>

configuration
>auxiliaries

and then <ENTER>.

>auxiliaries
dialog: english

This menu item has a colon, indicating that here the parameters can

be selected from a list. Press <SELECT> several times in order to

view the various selections and get used to this type of dialog.

>auxiliaries
dialog: deutsch

With <ENTER> you can accept the suggestion

'dialog:

eng-

lish'

.

>auxiliaries
display contrast: 3

4

×

<ê>

<ENTER>

By pressing <ê> 4 times you reach the menu selection

'max. lift

way'

.

>auxiliaries
max. lift way 125

mm

Here the lowest allowable lift position for automatic and manual

operation can be set. For the 6.2041.430 sample rack with

6.2743.050 sample tubes, this limit value of 125 mm should not be

changed. Accept the value by pressing <ENTER>.

configuration
>rack definitions

QUIT

or

STOP

In order to put the 766 IC Sample Processor back into the initial

position, press <QUIT> or <STOP>.

******** counter 0/127
PUMP- ready

In the normal state, the method name and the sample counter

reading are displayed in the first line. The second line serves as

status line which displays the pump status and the changer status.