6.4 Diagnosis

766 IC Sample Processor

123

6.4.9

Beeper

•

Prepare instrument for diagnosis (see section 6.4.2).

•

If necessary, press <ê> several times until

diagnosis

>beeper test

•

<ENTER>

The beeper is turned on and off in an endless loop.

>beeper test

•

The test can be exited by pressing the <QUIT> or <STOP> keys.

diagnosis
>rack code test

6.4.10 Rack code recognition

This diagnosis step checks the functionality of the sensors with regard

to their ability to automatically recognize the attached rack.

•

Note the code (magnets) on the underside of all the racks being

used. The following diagram shows the magnet fixture on the under-

side of a rack.

Magnets inserted

In this example
The code is:

000110

Magnet fixture as viewed from underneath

•

Pick up the rack and lay it on its side.

•

Prepare the instrument for diagnosis (see section 6.4.2).

•

If necessary press <ê> several times until

diagnosis
>power on reset

•

<ENTER>

The sample changer runs through its start routine (initializing lift and rack positions). The ini-

tialization is important because it is vital for the following diagnosis step “rack code test“ that

the rack is in the basic position.

•

Prepare instrument for diagnosis (see section 6.4.2).

•

If necessary press <ê> several times until

diagnosis
>rack code test