background image

4 Operation

766 IC Sample Processor

70

4.4

Manual operation

4.4.1

Turning the sample rack / Positioning the samples

Using the <ç> and <è> keys the sample rack can be
turned one position to the left (in the counterclockwise direction)
or right (clockwise).

8

MOVE

manual operation
MOVE 1 : sample

sample,spec.1

8,

1

999

Position vessel /
Turn rack

With the MOVE command, a particular sample tube or other
vessel can be placed under the needle. With <SELECT> the
numerical rack position as well as the predefined current sam-
ple (SAMPLE command) or the special beakers 1-8 can be
chosen.

For security reasons turning the sample rack is only possible

when the lift is in or above the shift position.

4.4.2

Moving the lift

The keys <é> and <ê> allow upward and downward move-
ment of the lift. The lowest possible lift position is defined by the
configuration parameter

'max. lift way'

.

HOME

END

The <HOME> key runs the lift to the rest position (0 mm), i.e. to
the upper limit.
<END> runs the lift to the predefined work position (see sec-
tion 4.3.2
).

9

LIFT

manual operation
LIFT: 1 : rest mm

work,rinse,shift,

special,rest,

0

125 mm

Position the lift

With the LIFT command, the lift can be run to a given position.
In addition to selecting an exact position in mm (0 - 125 mm),
the <SELECT> key can select a predefined position (work po-
sition, rinse position, shift position, special position, rest position
= 0 mm).