4 manual operation – Metrohm 766 IC Sample Processor User Manual
Page 75

4 Operation
766 IC Sample Processor
70
4.4
Manual operation
4.4.1
Turning the sample rack / Positioning the samples
Using the <ç> and <è> keys the sample rack can be
turned one position to the left (in the counterclockwise direction)
or right (clockwise).
8
MOVE
manual operation
MOVE 1 : sample
sample,spec.1
…
8,
1
…
999
Position vessel /
Turn rack
With the MOVE command, a particular sample tube or other
vessel can be placed under the needle. With <SELECT> the
numerical rack position as well as the predefined current sam-
ple (SAMPLE command) or the special beakers 1-8 can be
chosen.
For security reasons turning the sample rack is only possible
when the lift is in or above the shift position.
4.4.2
Moving the lift
The keys <é> and <ê> allow upward and downward move-
ment of the lift. The lowest possible lift position is defined by the
configuration parameter
'max. lift way'
.
HOME
END
The <HOME> key runs the lift to the rest position (0 mm), i.e. to
the upper limit.
<END> runs the lift to the predefined work position (see sec-
tion 4.3.2).
9
LIFT
manual operation
LIFT: 1 : rest mm
work,rinse,shift,
special,rest,
0
…
125 mm
Position the lift
With the LIFT command, the lift can be run to a given position.
In addition to selecting an exact position in mm (0 - 125 mm),
the <SELECT> key can select a predefined position (work po-
sition, rinse position, shift position, special position, rest position
= 0 mm).