Amplifier programming, Timing logic programming, All cd series modules) – Banner MAXI-AMP Series User Manual

Page 5: Cd5 models), Switches #2 and #3, Switch #4, Switch #1

Advertising
background image

Amplifier Programming

(all CD Series modules)

The diagram at the left shows the location of switches 1-4,
and the table summarizes the settings required for each
response condition.

NOTE: An adhesive-backed mylar label is supplied. It may
be marked to indicate switch programming and then ap-
plied to the MAXI-AMP housing as a switch cover.

Factory settings are
shown at left. The
"underlined" settings in
table below are factory
settings.

OFF

ON

OFF

AMPLIFIER

PROGRAMMING

#1

#2

#3

#4

SWITCH

SWITCH

SWITCH

SWITCH

ON

O

F

F

1

2

3

4

Modulation freq. "A"

1.5 millisecond response

15 millisecond response

Light operate

Dark operate

ON

ON

OFF

Modulation freq. "B"

OFF

ON

ON

OFF

ON

ON

ON

ON

OFF

ON

OFF

ON

PROGRAMMING

TIMING LOGIC

#1

#2

#3

#4

#5

#6

#7

#8

#9

#10

10

ON

O

F

1

2

4

3

F

5

6

7

8

9

SWITCH

SWITCH

SWITCH

SWITCH

ON

ON

ON

ON

OFF

OFF

OFF

SWITCH

SWITCH

SWITCH

SWITCH

SWITCH

SWITCH

On/Off

On Delay

Off Delay

On and Off Delay

One-shot

Delayed One-shot

Limit

Repeat Cycle

AC Latch

DC Latch

Delay and Latch

Limit and Latch

ON

ON

OFF

ON

OFF

ON

OFF

OFF

OFF

OFF

ON

ON

OFF

OFF

OFF

ON

OFF

ON

OFF

OFF

OFF

OFF

OFF

ON

ON

ON

OFF

OFF

OFF

OFF

OFF

ON

ON

OFF

ON

OFF

ON

ON

OFF

ON

ON

ON

ON

ON

ON

OFF

ON

ON

ON

ON

ON

OFF

OFF

ON

ON

ON

ON

ON

OFF

OFF

OFF

ON

ON

N/C Output

N/O Output

.15 Sec. Max. Time

1.5 Sec. Max. Time

15 Sec. Max. Time

OFF

OFF

OFF

OFF

ON

ON

OFF

ON

OFF

ON

OFF

ON

5

A group of ten switches, located on the side of the module opposite the amplifier program
switches, is used to select the timing logic for the CD5 models.

Switches #1 through #7 are used to select the logic function. Switch #8 programs the output for
either NORMALLY OPEN or NORMALLY CLOSED operation. Switches #9 and #10
program the time range(s). There are three ranges: 10 to 150 milliseconds, 0.1 to 1.5 seconds,
and 1 to 15 seconds. The programmed range will be the same for both functions of a dual timing
mode (ON & OFF DELAY, DELAYED ONE-SHOT, and REPEAT CYCLE). However,
DELAY and HOLD times are independently adjustable within the selected range.

The diagram shows switch locations, and the
table summarizes the program switch posi-
tions.

tion frequencies. This makes it possible to use the two sensor pairs
in close proximity without optical crosstalk. Both amplifiers must
be set for the same response time (either 1.5 or 15 milliseconds) to
ensure freedom from crosstalk.

Switches #2 and #3

are used to program the amplifier response

time. The 15 millisecond setting allows SP12 Series sensors to
operate at their maximum excess gain.

Switch #4

is used to select LIGHT OPERATE or DARK OPER-

ATE. In the LIGHT OPERATE mode, the output will energize (in
ON/OFF or LATCH operation) or the timing function will initiate (in
DELAY, ONE-SHOT, or LIMIT operation) when the receiver
"sees" sufficient light (excess gain greater than 1X). In DARK
OPERATE, the output will energize or timing will begin when the
receiver is sufficiently dark (excess gain less than 1X).

Amplifier response conditions may be programmed via the group of four
switches located on one of the narrow sides of the MAXI-AMP™
module.

Switch #1

selects the modulation frequency of the amplifier and the

emitter light source. When two pairs of SP12 Series sensors are being
used in close proximity to each other, the modulation frequency switch
of their respective CD Series modules should be set to different modula-

Settings illustrated below are factory settings. Factory settings are "underlined" in the table.

Timing Logic Programming

(CD5 models)

Advertising